Skip to main content
. Author manuscript; available in PMC: 2008 Sep 18.
Published in final edited form as: Rapid Commun Mass Spectrom. 2008;22(2):75–83. doi: 10.1002/rcm.3331

Table 2.

Linearity, limit of detection (LoD) and limit of quantitation (LoQ) of the ESI-LC-MS/MS assay for hydroxy fatty acids, leukotrienes, resolvins and protectins. (LoD and LoQ are expressed as pg on-column).

Compound Equation Correlation
Coefficient
LoD (pg) LoQ (pg)
9-HODE y = 2.080x − 0.010 0.994 10 20
13-HODE y = 0.483x − 0.028 0.987 20 50
5-HEPE y = 0.511x + 0.021 0.998 20 50
18-HEPE y = 0.279x + 0.010 0.994 20 50
9-HEPE y = 0.321x − 0.016 0.996 20 50
8-HEPE y = 0.877x + 0.037 0.972 20 50
15-HEPE y = 0.459x + 0.016 0.999 20 50
12-HEPE y = 0.484x − 0.030 0.999 20 50
5-HEPE y = 0.564x + 0.050 0.986 20 50
9-HETE y = 0.169x + 0.043 0.983 20 50
8-HETE y = 2.688x + 0.073 0.997 10 20
11-HETE y = 4.623x − 0.046 0.999 10 20
15-HETE y = 1.083x − 0.013 0.998 10 20
12-HETE y = 0.472x − 0.007 0.991 20 50
LTB4 y = 2.127x + 0.031 0.991 10 20
17S-HDHA y = 1.404x + 0.044 0.992 10 20
RvE1 y = 1.949x + 0.208 0.972 20 50
PD1 y = 0.606x + 0.125 0.984 10 20
RvD1 y = 2.979x + 0.008 0.996 20 50