Skip to main content
. Author manuscript; available in PMC: 2008 Sep 22.
Published in final edited form as: Anal Chem. 2005 Oct 15;77(20):6719–6728. doi: 10.1021/ac050987t

Table 3.

Determination of Fluoride in Drinking Water Samples Using Al[OEP]/ETH-7075 Based Fluoride Optical Sensitive Membrane, Film-6, and LaF3 ISE in 100 mM β-Alanine buffer, pH 3.60, 1 mM CDTA.

Al[OEP]/ETH-7075 LaF3-ISE

Sample ppm c.v., %a ppm c.v., %a
1 0.73 2.5 0.70 0.42
2 0.51 2.1 0.54 0.42
3 0.46 3.2 0.47 0.64
4 0.36 2.9 0.39 0.75
5 0.12 3.1 0.16 0.75
a