Table 1.
Implant | Indication for Compress® implant | Location of Compress® implant | Reason for retrieval | Age at time of retrieval (years) | Compression force (lb) | Duration of implantation (years) |
---|---|---|---|---|---|---|
1 | Failed oncology implant | Distal femur | Infection | 50 | 400 | 5.3 |
2 | Osteosarcoma | Distal femur | Fracture | 17 | 800 | 1.7 |
3 | Osteosarcoma | Distal femur | Infection | 10 | 400 | 1.1 |
4 | Failed total hip replacement | Distal femur | Infection | 64 | 400 | 1.3 |
5 | Knee arthrodesis takedown | Distal femur | Infection | 43 | 800 | 3.1 |
6 | Failed oncology implant | Proximal tibia | Infection | 43 | 600 | 0.4 |
7 | Septic prosthetic arthritis | Distal femur | Infection | 21 | 800 | 0.8 |
8 | Recurrent chondrosarcoma | Proximal femur | Tumor recurrence | 59 | 800 | 2.6 |
9 | Failed oncology implant | Distal femur | Infection | 65 | 800 | 6.5 |
10 | Osteosarcoma | Proximal humerus | Tumor recurrence | 15 | 600 | 1.2 |
11 | Osteosarcoma | Proximal tibia | Infection | 15 | 400 | 3.2 |
12 | Osteosarcoma | Distal femur | Infection | 44 | 400 | 12.2 |
Average (range) | 37 (10–65) | 600 (400–800) | 3.3 (0.4–12.2) |