Skip to main content
. Author manuscript; available in PMC: 2009 Sep 1.
Published in final edited form as: Acad Radiol. 2008 Sep;15(9):1099–1110. doi: 10.1016/j.acra.2008.04.004

Table 1.

Results from the Geometric Accuracy Test from T1-weighted Images. ACR specifies that lengths should be with ± 2 mm of actual length shown in first column.

Length (mm) N Mean Standard Deviation Min Max Coefficient Variation
P1 148 259 146.91 1.27 140.00 149.50 0.87
P2 190 260 189.84 1.06 182.00 192.00 0.56
P3 190 260 189.94 0.78 186.00 192.00 0.41
P4 190 260 189.92 0.91 185.00 192.00 0.48
P5 190 260 189.85 1.05 180.10 193.00 0.55
P6 190 260 189.71 0.93 186.40 192.40 0.49
P7 190 260 189.78 0.83 186.90 192.00 0.44