Skip to main content
. Author manuscript; available in PMC: 2008 Oct 8.
Published in final edited form as: Anal Chem. 2006 Nov 1;78(21):7461–7466. doi: 10.1021/ac060995p

Table 1.

Reference emission ratios, sensitivity, linearity, and signal reproducibility of 40:60% (v:v) AHAP3/ETMOS-TMOS NO-releasing sensors

Sensor 580 / 640 nm Emission Ratioa Δ Emission Ratio / Δ pH R2 Average standard deviation Resolvable pH Shift
1 0.850 0.196 0.986 0.00392 0.040
2 0.589 0.113 0.999 0.00203 0.036
3 0.707 0.116 0.999 0.00173 0.030
4 0.502 0.075 0.999 0.00138 0.037
5 0.581 0.079 0.998 0.00173 0.044
a

pH=7.4