Table 3.
Factor | Linked with revision event? |
---|---|
Tibial polyethylene shelf age | Yes (p < 0.01) |
Release of lateral collateral ligament and popliteus | Yes (p < 0.01) |
Patient age | No |
Patient weight | No |
Patient gender | No |
Preoperative tibiofemoral valgus angle | No |
Postoperative tibiofemoral valgus angle | No |
Use of titanium baseplate | No |
Use of sagittally flat insert | No |
Release/pie-crusting of iliotibial band | No |
Recession/release of posterior cruciate ligament | No |
Cox proportional hazards regression analysis was used to identify if the 11 listed factors were associated (p < 0.05) with an event of revision as a result of tibial polyethylene wear, osteolysis, or instability (the time variable of the analysis for each of the 100 knees was the number of years that had elapsed from the date of surgery to the date that the implants were last known to be in situ).