. 2008 Aug 19;466(11):2644–2649. doi: 10.1007/s11999-008-0436-8

Table 3.

Results of regression analysis testing for associations between 11 factors and revision events

Factor Linked with revision event?
Tibial polyethylene shelf age Yes (p < 0.01)
Release of lateral collateral ligament and popliteus Yes (p < 0.01)
Patient age No
Patient weight No
Patient gender No
Preoperative tibiofemoral valgus angle No
Postoperative tibiofemoral valgus angle No
Use of titanium baseplate No
Use of sagittally flat insert No
Release/pie-crusting of iliotibial band No
Recession/release of posterior cruciate ligament No

Cox proportional hazards regression analysis was used to identify if the 11 listed factors were associated (p < 0.05) with an event of revision as a result of tibial polyethylene wear, osteolysis, or instability (the time variable of the analysis for each of the 100 knees was the number of years that had elapsed from the date of surgery to the date that the implants were last known to be in situ).