. 2008 Aug 13;46(10):3319–3324. doi: 10.1128/JCM.00498-08

TABLE 4.

Performance evaluation of the DRW and Determine HBsAg rapid tests in China and Guinea using EIA as the gold standarda

Country Test Sensitivity
Specificity
No. positive/true positivesb % (95% CI) No. negative/true negativesc % (95% CI)
China DRW 185/186 99.46 (98.41-100) 481/485 99.18 (98.37-99.98)
Determine 184/186 98.92 (97.44-100) 485/485 100 (100-100)
Guinea DRW 174/180 96.66 (94.04-99.29) 399/399 100 (100-100)
Determine 170/180 94.44 (91.10-97.79) 399/399 100 (100-100)
a

The sensitivities and specificities of both tests were calculated relative to the results of the Hepanostika HBsAg Ultra EIA (BioMerieux, Lyon, France).

b

Number of samples positive by the DRW or Determine HBsAg rapid test/number of samples positive by two EIAs. The 186 samples from China comprise 15 EIA-positive samples from 500 randomly selected blood donor samples and 171 HBsAg-positive clinical samples. The 180 samples from Guinea comprise 92 EIA-positive samples from 491 randomly selected blood donor samples and 88 stored HBsAg-positive samples.

c

Number of samples negative by the DRW or Determine HBsAg rapid test/number of samples negative by two EIAs.