Published in final edited form as: Invest Ophthalmol Vis Sci. 2008 May 30;49(10):4290–4296. doi: 10.1167/iovs.08-2113

Table 5.

Correlation of Macular Subfield Thickness Measurements to Measures of Repeatability (CRW or CVW) of Study Eyes Included for Analysis

Pearson Correlation Coefficient (95% CI) P Pearson Correlation Coefficient (95% CI) P
Stratus OCT* central 0.32 (-0.03-0.60) 0.070 Cirrus HD-OCT* central -0.02 (-0.36-0.33) 0.927
Stratus OCT inner superior 0.36 (0.01-0.62) 0.043 Cirrus HD-OCT inner superior 0.27 (-0.09-0.56) 0.136
Stratus OCT inner nasal 0.51 (0.20-0.72) 0.003 Cirrus HD-OCT inner nasal 0.44 (0.11-0.68) 0.011
Stratus OCT inner inferior 0.31 (-0.04-0.59) 0.081 Cirrus HD-OCT inner inferior 0.47 (0.16-0.70) 0.005
Stratus OCT inner temporal 0.26 (-0.09-0.56) 0.137 Cirrus HD-OCT inner temporal 0.05 (-0.30-0.39) 0.772
Stratus OCT outer superior 0.56 (0.26-0.76) 0.001 Cirrus HD-OCT outer superior 0.25 (-0.10-0.55) 0.160
Stratus OCT outer nasal 0.56 (0.26-0.76) 0.001 Cirrus HD-OCT outer nasal 0.58 (0.29-0.77) < 0.001
Stratus OCT outer inferior 0.39 (0.05-0.64) 0.026 Cirrus HD-OCT outer inferior 0.29 (-0.06-0.58) 0.099
Stratus OCT outer temporal 0.17 (-0.18-0.49) 0.334 Cirrus HD-OCT outer temporal 0.00 (-0.34-0.34) 0.996

n = 33. Statistically significant correlation coefficients (P < 0.05) are in bold.

*

Carl Zeiss Meditec, Inc., Dublin, CA.