Skip to main content
. 2008 Sep 17;46(11):3759–3765. doi: 10.1128/JCM.01022-08

TABLE 1.

Overall numbers of screening results comparing the Paracheck Pf and the present amperometric immunosensor method

Microscopy resulta (no. of samples) No. of samples with Paracheck Pf test result of:
No. of positive samples detected by Paracheck Pf at:
No. of samples with amperometric immunosensor result of:
No. of positive samples detected by amperometric immunosensor at:
Pf −ve Pf +ve Unclear 1:10b 1:50b Pf −ve Pf +ve Unclear 1:10b 1:50b
Pf +ve (43) 5 34 4 12 0 2 41 0 41 36
Pf −ve (70) 57 8 5 66 1 3
a

Pf +ve, P. falciparum positive; Pf −ve, P. falciparum negative.

b

Sample dilution.