Table 2. OAI MR QA acquisition parameters.

Options that were selected to reduce artifacts or improve the accuracy and precision of quantitative measurements include no phase or readout partial Fourier.

Monthly OAI Monthly ACR Annual ACR Additional
2D TSE IW 2D TSE IW 3D DESS 2D MSME SE 2D TSE IW* 2D TSE IW 2D SE T1W 2D SE PD/T2W
Plane Sagittal Axial Sagittal Sagittal Sagittal Axial Axial Axial
Matrix (phase) 307 307 307 269 555 555 256 256
Matrix (freq) 384 384 384 384 704 704 256 256
No. of slices 7 7 160 21 1 11 11 11
FOV (mm) 140 140 140 120 250 250 250 250
Slice thickness (mm) 3 3 0.7 3 20 3 5 5
Skip (mm) 6 6 0 0.5 10 7 5 5
Flip Angle (deg) 180 180 25 90 180 180 90 90
TE/TI (ms) 29 29 4.7 10, 20, 30, 40, 50, 60, 70 29 29 20 20 / 80
TR (ms) 1750 1750 16.3 2700 1000 1750 500 2000
BW (Hz/pixel) 352 352 183 250 355 355 300 300
NAV (NEX) 1 1 1 1 2 2 1 1
Echo train length 5 5 1 1 5 5 1 1
Phase Encode Axis A/P A/P A/P A/P A/P A/P A/P A/P
Slice Partial Fourier 1 1 0.75 0.75 1 1 1 1
X-Resolution (mm) 0.365 0.365 0.365 0.313 0.355 0.355 0.977 0.977
Y-Resolution (mm) 0.456 0.456 0.456 0.446 0.450 0.450 0.977 0.977
Scan time (min) 1.8 1.8 10.5 10.6 4.5 3.3 2.8 8.5
*

The sagittal TSE IW ACR acquisition initially used 256×256 in plane resolution which resulted in pixel dimension 0.98mm×0.98mm.