Published in final edited form as: Ultrasound Med Biol. 2008 May 15;34(10):1678–1687. doi: 10.1016/j.ultrasmedbio.2008.03.001

Table 2.

Ultrasonic exposure parameters and threshold results. MI is the estimated equivalent on-screen Mechanical Index. The sources include laboratory transducers (LAB), and diagnostic ultrasound probes from the GE Vingmed system (FPA) and from the Toshiba system (PVN) (see text).

Frequency Exposure Duration Position Spacing Threshold Threshold Threshold

Source MHz min mm MPa MI MPa/MHz
LAB 1.0 1.0 1 - 0.67 0.67 0.67
LAB 1.5 1.5 2 2.6 0.95 0.78 0.63
FPA 2.5 1.5 1 - 0.76 0.62 0.51
LAB 2.25 2.3 3 1.8 1.3 0.87 0.58
PVN-375AT 2.5 1 - 1.22 0.76 0.48
FPA 2.5 3.2 1 - 1.71 0.96 0.53
LAB 3.5 3.5 4 1.1 2.34 1.25 0.67
FPA 10 5.0 1 - - - -
LAB 5.0 5.0 6 0.8 2.78 1.24 0.56
FPA 10 7.4 1 - - - -
LAB 7.5 7.5 9 0.5 5.93 2.16 0.79