Skip to main content
. Author manuscript; available in PMC: 2008 Nov 24.
Published in final edited form as: J Magn Reson Imaging. 2007 Nov;26(5):1296–1302. doi: 10.1002/jmri.21141

Table 1.

Adverse Events Involving RF Burns in MRI Scanners Reported to the FDA, 2005-2006*

Date of event MRI system Field (T) Description Leadsa Coilsb Borec
11/17/06 GE 1.5 First degree burns/blisters, both arms, ctl coil N Y N
11/17/06 GE 1.5 Burn/blisters, arm N N Y
11/16/06 Philips 1.5 Burn/blister, left forearm Y N N
11/2/06 GE 1.5 Burn/blister, arm N N N
11/1/06 Philips 1.5 Burn/blister, arm N N N
10/16/06 Philips 1.5 Burns/blisters, both thighs N N N
9/15/06 GE 1.5 Second degree burn, right elbow N N N
8/1/06 Hitachi 1.5 Burn, neck (ctl receive coil) N Y N
7/21/06 FONAR 0.6 Third degree burns, back (no padding) Y N N
7/14/06 Philips 1.5 Burns/blisters, hip N N N
6/5/06 Philips 1.5 Burns/blisters, inner thighs N N N
6/9/06 Philips 1.5 Second degree burns/blisters, breast N N N
5/26/06 Philips 1.5 Burns/blisters, elbow, leg N N N
4/29/06 Philips 1.5 Burns/blisters legs (calf) N N N
3/17/06 GE 1.5 Second degree burn, elbow N N Y
3/16/06 GE 1.5 Burn/blisters, hand (breast exam) injury to two patients N N N
3/5/06 Philips 1.5 Second degree burns/blisters, inner thighs N N N
2/23/06 Philips 1.5 Second degree burns, shoulder N N N
2/13/06 GE 1.5 Burns, blisters, elbows N N Y
1/24/06 Siemens 1.5 Burn/blister abdomen N Y N
12/20/05 GE 1.5 Third degree burns, hands (at isocenter) Y N N
12/19/05 GE 1.5 Burn/blisters, hand (breast exam) Y N N
12/11/05 GE 1.5 Burn/blister, chest N N N
12/10/05 GE 1.5 Second degree burns, arm, biceps N N N
11/1/05 Siemens 1.5 Burn/blisters on calves N N N
10/26/05 Philips 3 Second degree burn, shoulder; first degree burn, knee N N N
10/13/05 GE 1.5 Second degree burn lower neck Y N N
9/30/05 GE 1.5 Second degree RF burn, elbow N N N
9/23/05 GE 1.5 Second degree burns (torso array coil) N Y N
9/21/05 Siemens 1.5 Burns, fingers and thighs touching N N N
9/14/05 GE 1.5 Burn/blister, elbow N N N
8/27/05 GE 1.5 Burn/blisters, hip (8 ch body array) N Y N
8/23/05 GE 1.5 Burn/blisters, hip (8 ch body array) N Y Y
8/16/05 GE 1.5 Burn/blisters, hip (8 ch body array) N Y N
8/15/05 Philips 1.5 Burns/blisters, elbows (head MRI) N N N
7/22/05 GE 1.5 Burn, breast (lumbar MRI) N Y N
6/15/05 GE Burns/blistering on wrist (wrist coil) N Y N
4/20/05 GE Burns/blisters, C-spine exam N Y N
3/17/05 GE 1.5 Burns/blisters shoulder N N N
3/4/05 Philips 1.5 Burns/blister, arm (coil contact) N N Y
2/14/05 GE 1.5 Burns/blisters shoulder, elbow (shoulder array) N Y N
2/14/05 GE 1.5 Second degree burns, both legs (torso array) Y Y N
2/7/05 Siemens 1.5 Burn/blister arm N N Y
*

Search term “magnetic resonance,” and screen for burns. This data is reported voluntarily, and does not represent the true incidence or distribution of such incidents (9).

a

Report links injury to leads, including coil cables, or metal (Y = yes; N = No).

b

Report notes use of surface coils/arrays during scan.

c

Report notes injury was associated with bore contact.

ctl = cervical-spine/thoracic/lumbar, ch = channel.