TABLE VI.

Peak NEC Rates in KCPS at the Peak Activity in mCi for all Degrees of Collimation

Mode: 2D 2.1D 2.3D 2.5D 2.7D 2.9D 3D
20cm 203 @60* 164 @50 145 @40 123 @25 102 @17 72 @10 52 @6
27cm 107 @60* 81 @50 71 @40 60 @25 47 @17 32 @10 24 @6
35cm 45 @60* 36 @50 28 @40 23 @25 18 @17 12 @10 9 @6
*

this was the highest activity simulated, but still below the peak