TABLE VII.

Simulated NEC Rates in KCPS at 10.1, 7.0, and 5.5 mCi for the 2.5-D and 2.7-D Partial Collimation Modes in Comparison to 2-D and 3-D Rates for all Phantom Diametes. Also See Fig. 12 for Graphic Representations

graphic file with name nihms-75021-t0015.jpg 10.1 mCi 7.0 mCi 5.5 mCi
20 cm 27 cm 35 cm 20 cm 27 cm 35 cm 20 cm 27 cm 35 cm
2D 61.9 32.9 14.4 44.4 23.7 10.4 35.5 18.9 8.3
2.5D 89.0 40.1 17.5 69.9 34.9 14.0 58.4 29.3 11.8
2.7D 92.8 43.0 16.4 78.5 36.6 14.2 68.1 31.9 12.4
3D 47.7 22.2 8.3 51.5 23.8 8.7 51.5 23.6 8.6