. 2008 Oct 15;15(12):1834–1838. doi: 10.1128/CVI.00238-08

TABLE 3.

Cohorts of samples tested by Enfer multiplex immunoassay and the Anigen lateral-flow kit and individual responses to ESAT-6, CFP-10, and MPB83 antigens

Test TB-positive animals (no.) Sensitivity (%) TB-negative animals (no.) Specificity (%)
Enfer multiplex immunoassay 522 93.1 1,489 98.4
ESAT-6 522 40.6a 1,489 86.6a
CFP-10 522 82.6a 1,489 69.7a
MPB83 522 78.5a 1,489 99.1a
Anigen lateral flow 214 83.6b 79 83.0c
a

Data were extracted from the multiplex assay for these individual antigens.

b

The sensitivity of the Anigen lateral-flow kit is 90.0% compared to that of culture-confirmed positives, and it is 85.1% compared to that of skin test-confirmed positives according to the manufacturer's instructions.

c

The specificity of the Anigen lateral-flow kit is 98.6% compared to that of skin test-confirmed negative animals according to the manufacturer's instructions.