TABLE 2.

Intra-assay and total assay precision results of each CLIA

Reagent and sample Grand mean (s/co) SDa (%CV)b
Intra-assay Total
Elecsys
    NCc 0.134 0.011 (7.8) 0.014 (10.2)
    PCd 9.58 0.451 (4.7) 0.543 (5.7)
Architect
    NC 0.074 0.007 (9.1) 0.006 (8.3)
    PC 3.21 0.115 (3.6) 0.129 (4.0)
Vitros
    NC 0.232 0.015 (6.5) 0.023 (9.9)
    PC 5.53 0.173 (3.1) 0.191 (3.5)
Access
    NC 0.097 0.005 (5.5) 0.007 (7.2)
    PC 3.06 0.084 (2.8) 0.138 (4.5)
a

SD, standard deviation of s/co ratio.

b

CV, coefficient of variation of s/co ratio.

c

NC, negative control material.

d

PC, positive control material.