Table I.

Susceptibility artifacts averaged over four measurements in each condition. A summary of susceptibility artifacts measured from three electrode types (Ag/Agcl, Ag/AgCl coated carbon, and carbon wire) in two MRI imagers with different field strengths (1.5T and 3T) are shown.

Scanner 1.5T 3T
Electrode (mm) (mm)
Ag/AgCl (9 mm diameter, Compumedics) 6.1 9.4
Ag/AgCl coated carbon (10 mm diameter, MVAP Medical Supplies, 30222) 1.9 4.2
Carbon wire (9 mm diameter loop) 0.0 0.9