Skip to main content
. Author manuscript; available in PMC: 2008 Dec 6.
Published in final edited form as: J Biomed Opt. 2008;13(4):044038. doi: 10.1117/1.2952295

Table 1.

Noise contributions of the photodiode dark current, the integrated op−amp input noise, the free op−amp input noise, and Johnson noise from feedback resistors. The addition of the noises in quadrature predicts the total instrument noise, 5.67 μV. After instrument noise, the RMS noise contributions of illumination and physiological sources are tabulated in−vitro (lobster) and in−vivo (rat) and compared to the peak amplitudes of the stimulus−evoked optical changes. The standard errors for these measurements are large due to variability of signal strength and illumination intensity from trial to trial, but values give the reader a general idea of each source’s noise generation.

Noise source RMS Contribution (Δf = 500Hz), μV Quantity in Circuit
PD Op amp (UDT−555UV), Input Noise Voltage 0.45 1
PD (UDT−555UV/LN), Dark Current Noise 0.91 1
Op amp (TL074), Input Noise Voltage 4.00 2
1 Mohm Resistor Johnson Noise 2.88 4
Total Expected Instrument Noise in Quadrature 5.67 μV
   Measured Instrument Noise 4.53 μV
  Noise Source LED LD
Lobster, In−vitro μV dI/I × 104 μV dI/I × 104
Emitter (shot, speckle) RMS Noise 114.1 0.70 190.0 9.55
    Peak Signal Size In−vitro 128.6 ± 34.2 1.15 ± 0.34 48.1 ± 12.7 1.83 ± 0.46
  Rat, In−vivo μV dI/I × 104 μV dI/I × 104
Emitter (shot, speckle) RMS Noise 77.4 ± 22.0 0.64 ± 0.17 225.5 ± 111.5 3.51 ± 0.91
Cardiac RMS Noise (3Hz) 215.0 ± 72.2 1.82 ± 0.60 106.6 ± 48.5 1.72 ± 0.64
Respiration RMS Noise (1Hz) 182.1 ± 3.6 1.55 ± 0.02 56.5 ± 6.0 1.08 ± 0.08
Mayer Waves RMS Noise (0.0 to 0.4Hz) 616.8 ± 401.0 5.22 ± 3.37 550.3 ± 482.2 8.26 ± 7.19
  Total Noise Measured In−vivo 824.0 ± 325.8 6.98 ± 2.72 680.7 ± 446.0 10.7 ± 6.2
    Peak Signal Size In−vivo 381.7 ± 262.1 3.35 ± 2.19 162.3 ± 64.7 2.97 ± 0.66