Skip to main content
. Author manuscript; available in PMC: 2008 Dec 8.
Published in final edited form as: Nat Protoc. 2008;3(9):1422–1434. doi: 10.1038/nprot.2008.138

TABLE 2.

Troubleshooting table.

Potential problem(s) Possible source(s) and solution(s)
From the beginning of the experiment, the uncalibrated
volume signal is in the normal range (in mice 4–7;
in rats 5–10 RVUs), pressure signal and dP/dt is in the
lower range, the heart rate is very low (see Table 1
for normal values)
The temperature of the animal may have dropped or the anesthetic is overdosed. Check
the temperature of the animal and the temperature control unit; also check the depth
of the anesthesia and make sure that the anesthetic was not overdosed. If so, in case
of inhalation anesthesia, decrease the concentration by 1–1.5%, or for a short period,
use just 100% oxygen. Make adjustments and allow 10–15 min for stabilization. In case
of injectable anesthetic overdose, i.v. fluid injection and 100% oxygen can help recovery.
Make sure that animals are always intubated if injectable anesthesia is performed and
trachea tubes are not blocked by mucus. If so, insert a small plastic tube (P10 in mice) to
the tube connected to a 10–20 ml syringe and suck out obstructing mucus
From the beginning of the experiment, pressure
signal and dP/dt is low
Animal may have lost too much blood during the surgery and/or fluid via evaporation
through the surgical surface. Anesthesia/analgesia may also be overdosed or not
sufficient. Use battery-operated electrocautery during the surgery to minimize blood loss
and always moisturize the surgical surface with physiological saline to decrease drying out
and evaporation. If the blood/fluid loss is significant, slowly inject/infuse physiological
saline solution into the jugular or femoral veins; in more severe cases, inject albumin-
containing solution. If the abdominal cavity is intact, saline solution can also be injected
i.p. Check the anesthesia/analgesia depth. Paradoxically, if animals feel pain and distress,
it can also lead to hemodynamic instability; if so, adjust the drug concentrations
From the beginning of the experiment, pressure signal
and dP/dt is normal, but volume signal is very low
Most likely, the catheter is not in the right position. Try to improve the volume signal by
gently readjusting the position of the catheter
The pressure and/or volume signal is noisy, regular
50 Hz (Europe) to 60 Hz (USA) noise pattern
The most likely cause is electric interference, which can originate from pumps,
electrocautery devices, ungrounded operating tables, ventilators/respirators, fluorescent
lamps, transformers, fans and electric warming blankets. Try to isolate any sources
of electrical interference by moving them away from the catheter and by sequentially
unplugging the possible interfering appliances. If the noise is coming from the power
supply, sometimes just unplugging the MPVS system and plugging into different outlet,
preferably with stabilized power supply for the laboratory equipment, can eliminate the
problem. If the noise problem from the electrical network cannot be resolved for any
reasons, the built-in 50- and 60-Hz filter in PowerLab can sometimes improve results
The pressure and/or volume signal is noisy and shows
irregular pattern
If all of the above fail and the noise is irregular, check if it disappears with another
catheter. If so, the first catheter is most likely damaged. Check catheter for damage
under the microscope, and if the coating is disrupted, clean and send it immediately to
the manufacturer for further evaluation and repair. Always have at least one or two
backup catheters before starting a study
The pressure sensor shows an excessive amount of
drift, and/or it is not possible to ‘zero balance’ the
pressure transducer output
Fluid may have gotten inside the catheter, or the pressure sensor diaphragm may be
cracked or broken. It is possible to check the functionality of the pressure sensor by
connecting an ohmmeter (resistance gauge) across the pins of the pressure sensor
connector. The bridge on the Millar sensor should have symmetrical input and output
impedances of approximately 1,000 ohms. If either reading is dramatically different
from 1,000 ohms, there are chances that the pressure sensor is broken because of the
above-mentioned reasons. Please contact Millar Instruments' customer support
The pressure sensor appears to be functional, but the
readings are off by approximately 25 or 100 mm Hg
The pressure calibration in the data acquisition hardware is incorrect. Recalibrate the
pressure signal in the data acquisition software according to the manufacturer's
instructions. Briefly, the pressure control side of the MPVS-300/400 has three
calibration settings. The settings are 0, 25 and 100 mm Hg. The standardized pressure
output from the MPVS-300/400 is 1 V per 100 mm Hg. Therefore, the 0 mm Hg
calibration setting corresponds to 0 V output, 25 mm Hg corresponds to 250 mV (0.25 V)
output, and 100 mm Hg corresponds to 1 V output. In the case where the pressure
readings are off by exactly 25 or 100 mm Hg, the wrong units label has been assigned to
the voltage output coming out of the MPVS-300/400 for a particular calibration setting
(e.g., the 25 mm Hg units label was assigned to 0 V out rather than 0 mm Hg units label)
The zero balanced pressure sensor in saline or distilled
water drifts a few mm Hg after insertion into the
biological environment
The most likely reason is that the catheter was not properly prepared before inserting it
to the animal. Presoak the catheter tip in saline at body temperature for at least 30 min
before use (e.g., by inserting the tip (but not the connectors) through the tip of a
1-ml syringe containing physiological saline solution)
The volume/conductance signal output from the
MPVS-300/400 appears to drift
Electronic hardware has some drift present in the signal as a result of the electronic
components heating up. To minimize conductance signal drift from the MPVS-300/400,
turn it on and let it warm up at least 30 min before calibrating the output and using the
catheter to collect PV data
The calibrated volume signal is larger than expected Following the cuvette calibration, the volume signal (converted from RVUs to uLs) is
larger than expected. The reason for this is that the parallel volume contribution of the
myocardium (parallel conductance, which refers to the conductivity of the heart muscle
that surrounds the left ventricular blood pool) has not been taken into account. Perform
hypertonic saline calibration to correct for Vp
The conductance/volume signal is out of range In larger animals with dilated cardiomyopathy, the conductance signal may go out of
the range. Measure the ventricular length of the animal in question and make sure that
the catheter being used has electrode spacing that matches this length (e.g., a 6-mm
signal electrode spacing is appropriate for small-sized rats and a 9-mm signal electrode
spacing is appropriate for larger rats). In some cases, custom catheter design may be
required; contact the manufacturer with such request
The volume/conductance outputs from the cuvette
calibration are not consistent
Make sure that the cuvette with fresh heparinized blood (to avoid clotting) is at body
temperature. Always properly clean the cuvette between uses. Center the catheter tip
within the cuvette and submerge it (all four electrodes should be submerged) in the
blood, and hold it as steady as possible for 10–20 s in each well (normally using the first
1–5 wells is sufficient)