Skip to main content
. 2008 Oct 16;467(1):84–93. doi: 10.1007/s11999-008-0541-8

Table 3.

Mean depth of cement penetration measured at the side, chamfer, and dome of the femoral head for each implant design*

Implant Dome depth (mm) Chamfer depth (mm) Side depth (mm)
BHR 13.9 ± 5.5 13.1 ± 3.7 11.0 ± 4.7
ASR® 1.7 ± 1.1 1.7 ± 0.9 1.0 ± 0.3
Conserve Plus® 2.4 ± 1.2 2.6 ± 1.0 1.4 ± 0.3
DuROM® 2.4 ± 0.8 2.8 ± 0.9 1.7 ± 0.4
ReCAP® N/A N/A N/A

* Mean depth was measured perpendicularly at 1-mm intervals along the periphery of the femoral head for each section and reported as a mean of the average for each femoral head ± standard deviation; statistical difference between BHR and all other groups (analysis of variance, p < 0.0001); N/A = not available as a result of semicircular shape of inner geometry of the prosthesis.