Skip to main content
. Author manuscript; available in PMC: 2009 May 1.
Published in final edited form as: Lab Chip. 2008 Mar 28;8(5):710–716. doi: 10.1039/b719301g

Fig. 2.

Fig. 2

(A) The results of a 2-D finite element model of a cell trap on a single electrode. A conductive media with a 100 V potential between the cell trap electrode and the ground electrode was used. (B) Enlarged view of the cell trap. Note that the majority of the electric field strength generated in the cell trap is greater than 10 kV cm−1.