Table 2007.05C.

HPLC-UV parameters

HPLC column Zorbax SB-C18, 5 μm, 3.0 mm id × 25 cm, Agilent
Mobile phase A 0.1% (v/v) trifluoroacetic acid-purified water
Mobile phase B 0.1% (v/v) trifluoroacetic acid-acetonitrile
Pump program
Time, min
%A
%B
Gradient curve
0 80 20 NAa
25 30 70 1b
30 0 100 1
31 80 20 1
40
80
20
NA
Detection Ultraviolet absorbance detector (390 nm)
Column temperature, °C 40
Flow rate Approximately 0.5 mL/minute; flow rate should be adjusted so that desired retention time will be obtained
Injected volume, μL 25
Injection time, min 40
a

NA = Not applicable.

b

A gradient curve of 1 indicates a linear transition.