Skip to main content
. Author manuscript; available in PMC: 2009 Nov 1.
Published in final edited form as: Anal Chem. 2008 Oct 3;80(21):8231–8238. doi: 10.1021/ac801317t

Figure 8.

Figure 8

A 250 μm id capillary was utilized with a valco union and a 100 μL Hamilton gas tight syringe to initially aspirate the sample, oil plug, and blank solution into the capillary. After the cartridge was formed it was screwed into a nanoport and pumped into the segmentation channel and analyzed using the desegmenting gated injector. Electropherograms recorded using a 525 V/cm field strength with 10 mM borate and 0.9 mM HPβ-CD as the separation buffer. Separation distances were 5 cm and injection widths were 200 ms.