Skip to main content
. Author manuscript; available in PMC: 2010 Jan 1.
Published in final edited form as: J Liq Chromatogr Relat Technol. 2009 Jan;32(2):261–272. doi: 10.1080/10826070802603351

Table 2.

Accuracy and Repeatability for Assaying NSC 737664 in Human Urine by HPLC with UV Detection.

NSC 737664
Concentration Added (μM) Accuracy (%) Repeatability (%)
25.0 108.5 12.4
10.0 101.4 7.8
5.0 99.5 7.9
2.50 95.7 10.6
1.00 105.3 8.8

Accuracy and repeatability of the assay were assessed from 9 different standard curves of NSC 737664 in human urine separately prepared and analyzed over an 8-week period.