Published in final edited form as: Med Phys. 2006 Apr;33(4):828–839. doi: 10.1118/1.2178449

Table I.

The fields (6 and 18 MV photons) used in the validation calculations are provided. A label is given for each test-field class, the jaw settings, the multileaf collimator (MLC) settings, the monitor units per field (MU), and the calculation method used. The calculation methods are indicated by letter abbreviations: Hybrid (HM), Analytic 1 (A1), Analytic 2 (A2), Pinnacle segmented multileaf collimator (SMLC) implemented in PINNACLE3 v6.2b (pinn v6.2b), and Pinnacle SMLC and sliding window (DMLC) implemented in PINNACLE3 v7.6c (pinn v7.6c).

Label Jaws (cm2) MLC setup MU Calculation methods
Leaf tips 10×24 Leaf tips on CAX 50 HM, A1, A2
Even teeth 10×24 odd retracted 100 HM, A1, A2
Odd teeth 10×24 even retracted 100 HM, A1, A2
Closed 10×24 closeda 999 HM, A1, A2
Closed 5×5 5×5 closed 100 HM, A1, A2
Closed 10×10 10×10 closed 100 HM, A1, A2
Closed 15×15 15×15 closed 100 HM, A1, A2
Closed 20×20 20×20 closed 100 HM, A1, A2
Open 10×10 open 100 HM, A1, A2
1 mm gap 13×10 1 mm sliding window 100 HM, A1, A2
10 mm gap 13×10 1 cm sliding window 100 HM, A1, A2
100 mm gap 13×10 10 cm sliding window 100 HM, A1, A2
Clinical DMLC IMRT fieldsb variable field specific Variable, 80-150 HM, A1, A2, pinn v7.6c
Clinical SMLC IMRT fieldsc variable field specific Variable, 80-150 HM, pinn v6.2b, pinn v7.6c
a

MLC leaves are blocking the field with the leaf tips positioned behind the jaws.

b

Total of 79 patient sliding window (DMLC) intensity modulated radiation therapy (IMRT) fields. 70 of these fields required split beams for delivery.

c

Total of 29 patient segmented multileaf collimator (SMLC) IMRT fields.