Table I.
The fields (6 and 18 MV photons) used in the validation calculations are provided. A label is given for each test-field class, the jaw settings, the multileaf collimator (MLC) settings, the monitor units per field (MU), and the calculation method used. The calculation methods are indicated by letter abbreviations: Hybrid (HM), Analytic 1 (A1), Analytic 2 (A2), Pinnacle segmented multileaf collimator (SMLC) implemented in PINNACLE3 v6.2b (pinn v6.2b), and Pinnacle SMLC and sliding window (DMLC) implemented in PINNACLE3 v7.6c (pinn v7.6c).
| Label | Jaws (cm2) | MLC setup | MU | Calculation methods |
|---|---|---|---|---|
| Leaf tips | 10×24 | Leaf tips on CAX | 50 | HM, A1, A2 |
| Even teeth | 10×24 | odd retracted | 100 | HM, A1, A2 |
| Odd teeth | 10×24 | even retracted | 100 | HM, A1, A2 |
| Closed | 10×24 | closeda | 999 | HM, A1, A2 |
| Closed 5×5 | 5×5 | closed | 100 | HM, A1, A2 |
| Closed 10×10 | 10×10 | closed | 100 | HM, A1, A2 |
| Closed 15×15 | 15×15 | closed | 100 | HM, A1, A2 |
| Closed 20×20 | 20×20 | closed | 100 | HM, A1, A2 |
| Open | 10×10 | open | 100 | HM, A1, A2 |
| 1 mm gap | 13×10 | 1 mm sliding window | 100 | HM, A1, A2 |
| 10 mm gap | 13×10 | 1 cm sliding window | 100 | HM, A1, A2 |
| 100 mm gap | 13×10 | 10 cm sliding window | 100 | HM, A1, A2 |
| Clinical DMLC IMRT fieldsb | variable | field specific | Variable, 80-150 | HM, A1, A2, pinn v7.6c |
| Clinical SMLC IMRT fieldsc | variable | field specific | Variable, 80-150 | HM, pinn v6.2b, pinn v7.6c |
MLC leaves are blocking the field with the leaf tips positioned behind the jaws.
Total of 79 patient sliding window (DMLC) intensity modulated radiation therapy (IMRT) fields. 70 of these fields required split beams for delivery.
Total of 29 patient segmented multileaf collimator (SMLC) IMRT fields.