Table III.

Averages of the average γ values and averages of the fraction of points with γ<1 with 2% /2 mm and 3% /3 mm criteria for the 79 dynamic multileaf collimator (DMLC) and 29 segmented multileaf collimator (SMLC) intensity modulated radiation therapy (IMRT) fields. The one-standard deviation for each quantity is given in parenthesis. For each IMRT type, one-way paired ANOVA tests indicated that the improved gamma results for the hybrid method is statistically significant.

2% /2 mm
3% /3 mm
IMRT type Method γ¯ % pass γ<1 γ¯ % pass γ<1
DMLC Hybrid 0.38 (±0.08) 92.2 (±3.7) 0.25 (±0.05) 99.0 (±0.9)
DMLC PINNACLE3 7.6c 0.4 (±0.1) 90.0 (±4.3) 0.27 (±0.08) 98.2 (±2.3)
DMLC Analytic 2 0.41 (±0.08) 89.0 (±4.2) 0.28 (±0.05) 97.0 (±2.2)
DMLC Analytic 1 0.85 (±0.2) 66.1 (±9.7) 0.56 (±0.14) 83.0 (±7.1)
SMLC Hybrid 0.27 (±0.1) 94.5 (±6.1) 0.19 (±0.08) 98.3 (±2.8)
SMLC PINNACLE3 7.6c 0.34 (±0.2) 92.2 (±7.7) 0.22 (±0.12) 97.1 (±4.3)
SMLC PINNACLE3 6.2b 0.41 (±0.22) 89.2 (±9.5) 0.27 (±0.14) 96.0 (±4.9)