Table 3.

Assessment of accuracy and precision from quality control samplesa

Nominal
(ng/mL)
GM
(ng/mL)
S.D.
(ng/mL)
DEV
(%)
BRP
(%)
WRP
(%)
n
3.00 2.86 0.161 -4.76 3.43 4.63 30
40 40.40 2.714 0.99 NA 7.31 30
80 79.54 4.079 -0.58 NA 5.19 30

Abbreviations: GM, grand mean; S.D., standard deviation; DEV, percent deviation from nominal value; WRP, within-run precision; BRP, between-run precision; n, total number of replicate observations within the validation runs. NA, not applicable, in this case it means no further between-run variance was observed.

a

Five samples at each concentration were run on six different occasions.