Table 4.
LROC Results Indicating Impact of Contrast on Detection for Frame av. The areas under the LROC curve indicate that RBI with RC significantly improves detection. However, at lower contrast levels RBI with AC_RC_SC provides the best results.
Area Under LROC Curve | ||||
---|---|---|---|---|
NC | AC | RC | AC_RC_SC | |
Contrast 6.5 | 0.32 ± 0.01 | 0.30 ± 0.09 | 0.48 ± 0.05 | 0.62 ± 0.06 |
Contrast 7.5 | 0.45 ± 0.05 | 0.57 ± 0.07 | 0.59 ± 0.04 | 0.81 ± 0.06 |
Contrast 9 | 0.64 ± 0.06 | 0.64 ± 0.07 | 0.79 ± 0.03 | 0.79 ± 0.06 |