Published in final edited form as: Phys Med Biol. 2008 Dec 19;54(2):433–445. doi: 10.1088/0031-9155/54/2/017

Table 2.

Average DOI error (standard deviation in mm) of the calibrated depths compared with known depths for all crystals for each of the eight detector arrays of the prototype scanner

Depth (mm) 2 10 18
Crystal Detector Crystal Detector Crystal Detector
Detector 1 0.39 0.97 0.66 0.76 0.43 1.09
Detector 2 0.52 1.29 0.49 0.35 0.73 0.89
Detector 3 0.70 1.21 0.50 0.37 0.39 0.73
Detector 4 0.60 1.26 0.57 0.53 0.31 0.98
Detector 5 0.47 1.36 0.47 0.46 0.23 0.99
Detector 6 0.20 0.67 0.21 0.18 0.37 0.42
Detector 7 0.44 0.71 0.60 0.71 0.37 0.79
Detector 8 0.43 1.43 0.81 1.35 0.53 1.02
Average 0.47 1.11 0.54 0.59 0.42 0.86