Table 1.

Assay performance data of the calibration samples for ABT-888 and M8 in human plasma.

Analyte Concentration (ng/mL) Accuracy (%) Intra-assay precision (%) Inter-assay precision (%)

ABT-888 10 100.4 4.5 -*
30 99.9 5.0 -*
100 100.6 3.3 -*
300 101.0 3.6 2.8
500 102.4 2.8 2.0
750 101.6 3.9 -*
1000 96.5 3.6 1.7

M8 10 101.4 5.7 -*
30 97.4 4.2 2.4
100 100.8 7.0 -*
300 102.3 5.5 -*
500 104.6 6.3 -*
750 104.4 12.1 -*
1000 98.3 11.3 -*

N=9; triplicate results, each in 3 separate runs, for each concentration.

*

The mean square of the within runs was greater than the mean square of the between runs, indicating that there was no significant additional variation due to the performance of the assay in different runs [9].