Table 1.

Troubleshooting guide for SPE of glycoprotein carbohydrates released by reductive β-elimination

MALDI analyte/matrix spots Cause Comments
Clear, translucent sample spots with no crystal definition Excess base (NaBH4 + NaOH) from β-elimination reaction SPE cartridge requires more thorough desalting with H2O prior to elution of glycans
Poor MALDI-FT-ICR-MS data
Dark sample spots with no crystal definition Excess base (NaBH4 + NaOH) from β-elimination reaction and HCl from base neutralization SPE requires more thorough desalting with H2O prior to elution of glycans
Poor MALDI-FT-ICR-MS data
Sample spots appear normal with good crystal definition Excess H2O desalting of SPE cartridge prior to elution, or too fast a washing rate Slow desalting rate (~1 ml/min)
Poor MALDI-FT-ICR-MS data Fine-tune volume of desalting wash to sample size

Note: In the presence of excess HCl, spots appeared normal with good crystal definition and no apparent effects on MALDI-FT-ICR-MS data.