. 2008 May 17;80(12):4741–4751. doi: 10.1021/ac702628q

Table 2. Assay Linearity and Precision for the WHO Spiked Samples across All Participating Laboratoriesa.

  log10 SD
detectable fold differenceb
   
cytokine/assay intra-assay interlaboratory intermatrix intra-assay interlaboratory intermatrix CV < 20% (%)c linearity r2
  IL-1β
BioSource 0.04 (0.05) 0.07 (0.19) 0.15 (0.20) 1.65 1.84 2.46 75.56 0.99 (0)
Endogen 0.03 (0.03) NAd 0.07 (0.08) 1.61 NA 1.84 96.67 0.98 (0.02)
Quantikine 0.01 (0.04) 0.13 (0.27) 0.14 (0.26) 1.5 2.35 2.39 86.67 1.00 (0.00)
QuantiGlo 0.01 (0.02) 0.08 (1.24) 0.08 (1.17) 1.45 1.9 1.89 98.33 1.00 (0.01)
Beadlyte Luminex 0.02 (0.02) 0.07 (0.29) 0.37 (0.28) 1.54 1.84 5.76 90 0.99 (0.01)
in-house Luminex 0.03 (0.03) 0.65 (0.08) 0.62 (0.07) 1.59 16.37 14.68 90 0.98 (0.03)
MSD Small Spot 0.03 (0.03) 0.05 (0.04) 0.07 (0.05) 1.56 1.68 1.88 96.67 0.99 (0.02)
  IL-6
BioSource 0.04 (0.05) 0.30 (0.31) 0.51 (0.18) 1.64 4.4 9.49 76.67 0.99 (0.01)
Endogen 0.03 (0.05) 0.48 (0.58) 0.66 (0.43) 1.59 8.48 16.82 76.67 0.98 (0.04)
Quantikine 0.02 (0.02) 0.07 (0.05) 0.21 (0.01) 1.54 1.86 3.12 91.67 1.00 (0.00)
QuantiGlo 0.03 (0.03) 0.08 (0.1) 0.28 (0.04) 1.56 1.89 4.06 93.33 0.99 (0.01)
Beadlyte Luminex 0.02 (0.03) 0.41 (0.53) 0.78 (0.33) 1.54 6.5 26.07 96.67 0.79 (0.51)
in-house Luminex 0.02 (0.03) 0.40 (0.16) 0.48 (0.18) 1.54 6.31 8.62 90 1.00 (0.01)
MSD Small Spot 0.03 (0.02) 0.04 (0.01) 0.21 (0.06) 1.56 1.63 3.1 100 0.99 (0.01)
a

Comparisons within (i) replicate samples (intra-assay), (ii) samples pooled from multiple laboratories (interlab), and (iii) samples pooled for the five different biological matrices (intermatrix). SD (log10 pg/mL) and linearity coefficient r2 are expressed as median (interquartile range).

b

The fold difference in cytokine concentration for which the assay would have 95% power to detect with P < 0.05.

c

The proportion (%) of measurements with an intra-assay CV <20%.

d

NA: not applicable (only one laboratory ran this assay).