Skip to main content
. Author manuscript; available in PMC: 2009 Feb 24.
Published in final edited form as: J Biomed Opt. 2008;13(6):064029. doi: 10.1117/1.3041497

Table 2.

Reconstruction results for phantom 1 (μa=0.23 cm-1 and μs=5.45cm1 using the semi-infinite model and the two-layer model with s fitted background values

Semi-Infinite Model
Two-Layer Model
Target Center Depth (C)/Second Layer Depth (D) Fitted Background Value [μa1μs1](cm1) Reconstructed Maximum μa (cm-1) Fitted Background Value [μa1μs1μa2μs2](cm1) Reconstructed Maximum μa (cm-1)
Phantom 1 anD silicon C 0.9 cm, D 1.5 cm [0.048 5.33] 0.158 (68%) [0.036 7.89 0.11 3.25] 0.201 (88%)
phantom 1 C 1.5 cm, D 2.1 cm [0.039 4.76] 0.169 (73%) [0.037 5.92 0.079 5.17] 0.206 (90%)
C 1.8 cm, D 2.4 cm [0.037 5.40] 0.191 (83%) [0.027 9.90 0.041 18.53] 0.198 (86%)
Phantom 1 and silicon C 0.9 cm, D 1.5 cm [0.054 4.79] 0.165 (72%) [0.020 5.62 0.083 20.21] 0.205 (89%)
phantom 2 C 1.5 cm, D 2.1 cm [0.037 5.42] 0.160 (70%) [0.027 5.53 0.051 5.59] 0.173 (75%)
C 1.8 cm, D 2.4 cm [0.033 5.57] 0.186 (81%) [0.018 7.71 0.051 8.07] 0.187 (81%)