TABLE 3.

Comparison of the endpoint detection limits of the multiplex and singleplex PCR assays

Virus and dilution Detection limits of assaya
Singleplex Multiplex
CMV
    10−1 23.94, 23.83 23.40, 23.73
    10−2 27.35, 26.96 26.79, 27.24
    10−3 32.20, 32.07 30.21, 30.34
    10−4 33.39, 33.28 33.20, 34.09
    10−5 36.27, − 37.43, −
EBV
    10−1 23.19, 23.22 23.36, 23.34
    10−2 26.37, 26.41 26.22, 26.25
    10−3 30.02, 29.58 30.00, 29.52
    10−4 32.92, 34.13 33.62, 32.89
    10−5 37.13, − 35.47, −
Adenovirus
    10−1 22.20, 22.63 22.40, 22.33
    10−2 26.21, 26.60 25.51, 25.60
    10−3 30.08, 29.64 29.08, 29.24
    10−4 32.99, 33.59 32.58, 32.59
    10−5 36.52, − 35.52, 35.09
a

Expressed as CT values obtained with samples tested in duplicate. −, no virus detected.