Table 4.
Quantification of short-term (20 min) human serum protein adsorption on bare and modified TiO2 sensors using OWLS. Standard deviations from the mean for 3 replicates are reported.
Substrate | Adsorbed Serum Mass (ng/cm2) |
---|---|
Bare TiO2 | 342 ± 21 |
PMP110 | 51 ± 7 |
PMP1-AMP* | 119 ± 40 |
PMP1-C* | 88 ± 30 |