Table 4.
Inter-, intra- and total-assay precision (ANOVA) for lower limit of quantification, upper limit of quantification and diluted over-curve samples for artemisinin in human plasma.
| Inter-assay CV (%) | Intra-assay CV (%) | Total-assay CV (%) | Accuracya (%) | |
|---|---|---|---|---|
| LLOQ: 1.03 ng/mL | 2.82 | 3.53 | 3.35 | 101 |
| ULOQ: 762 ng/mL | 8.02 | 2.16 | 4.57 | 99 |
| Over-curve: diluted to 572 ng/mL | 3.56 | 3.57 | 3.57 | 104 |
Mean accuracy over all 4 validation days.