Skip to main content
. Author manuscript; available in PMC: 2009 Mar 25.
Published in final edited form as: Proteomics Clin Appl. 2008;2(10-11):1378–1385. doi: 10.1002/prca.200780170

Table 1.

Recommended practices for clinical applications of protein profiling by MALDI TOF MS

Preanalytical Evaluate optimum patient preparation
Identify optimum procedures for specimen collection and processing
Analyze specimen stability
Develop criteria for specimen acceptability
Analytical Prepare calibrators for mass, resolution, and detector sensitivity
Use internal standards
Automate specimen preparation
Optimize methods to yield highest possible signals for peaks of interest
Identify sequences of peaks of interest
Develop calibration materials for components of interest
QC: prepare/identify at least two concentrations of control material
Evaluate reproducibility (precision)
Evaluate LOD and linearity
Evaluate reference intervals
Evaluate interferences such as hemolysis, lipemia, renal failure, acute-phase responses
Develop materials or programs for external comparison/proficiency testing of analyzers
Postanalytical Analyze each spectrum to identify peaks before applying diagnostic algorithms
Develop criteria for the acceptability of each spectrum based on peak characteristics
Use peaks rather than raw data as the basis for diagnostic analysis
Use caution in interpretation of peaks with m/z<1200
Select peaks with high intensities and sample stability for diagnosis
Select approximately equal numbers of peaks that increase and decrease in intensity as diagnostic discriminators
In developing a training set for diagnosis, careful clinical classification of patients is essential
Clinical validity depends on having a typical rather than highly selected population of patients
The number of training specimens should be at least ten times the number of measured values
Any clinical application should use a fixed training set and algorithm for analysis
Any analysis should provide a numerical value
Diagnostic performance should be evaluated with ROC curves to select cutoffs
A sensitivity analysis should be performed of the necessary precision for accurate diagnostic performance
There should be QC procedures for daily verification of software performance

Adapted from Hortin [26].