Skip to main content
. Author manuscript; available in PMC: 2009 Mar 29.
Published in final edited form as: Nat Protoc. 2006;1(6):2759–2762. doi: 10.1038/nprot.2006.399

TABLE 1.

Troubleshooting table.

Problem Possible reasons Solution
Color formation from Hcy is not observed at desired detection limit Hcy solution is not fresh Use sample containing reduced Hcy immediately
Low Hcy levels Run replicates; concentrations of methyl viologen dichloride hydrate and/or buffer might need adjustment (e.g., see ref. 15); time and/or temperature might need adjustment
Wrong buffer Buffer must be Tris; no substitutes are acceptable
Solution color fades rapidly Methyl viologen dichloride hydrate radical cation is quenched Sample should not be allowed to cool before visual inspection; do not move sample
Solution color formation is nonselective; observed with cysteine and/or other thiols pH is too high, resulting in nonselective formation of reducing disulfide radical cation Check buffer solution and check that the pH in the plasma sample is neutral using pH paper
Metal catalysis Avoid contact with metals; use deionized water
Solution turns dark greenish after 4 min Temperature is too high or heating period is too long Use milder heating
Foaming observed in vial Leak Make sure vials are tightly closed before heating