Table 1.

Required radiologic technique factor ranges

Parameter Operational mode
Fluoroscopy Image recording
Pulsed GI/GU Pulsed interventional Cardiac Non-cardiac
Pulse width (ms)
Maximum adults 8 8 8 50
Maximum pediatrics 5 5 5 20
Optimum pediatrics 4 4 4 10
Pulse rate (pulses/s) 1, 2, 4, 8 7.5, 15, 30 7.5, 15, 30 1, 2, 3, 4, 6
High voltage settings (kVp)
Optimum/standard filter 70 70 70 70
Minimum/standard filter 60 60 60 60
Maximum/standard filter 120 120 120 120
Optimum/spectral filter (filter thickness dependent; 50–0 kVp)
Tube current (mA)
Minimum 10 10 100 100
Maximum 100 100 1000 1000