Table 1.
Parameter | Operational mode | |||
---|---|---|---|---|
Fluoroscopy | Image recording | |||
Pulsed GI/GU | Pulsed interventional | Cardiac | Non-cardiac | |
Pulse width (ms) | ||||
Maximum adults | 8 | 8 | 8 | 50 |
Maximum pediatrics | 5 | 5 | 5 | 20 |
Optimum pediatrics | 4 | 4 | 4 | 10 |
Pulse rate (pulses/s) | 1, 2, 4, 8 | 7.5, 15, 30 | 7.5, 15, 30 | 1, 2, 3, 4, 6 |
High voltage settings (kVp) | ||||
Optimum/standard filter | 70 | 70 | 70 | 70 |
Minimum/standard filter | 60 | 60 | 60 | 60 |
Maximum/standard filter | 120 | 120 | 120 | 120 |
Optimum/spectral filter | (filter thickness dependent; 50–0 kVp) | |||
Tube current (mA) | ||||
Minimum | 10 | 10 | 100 | 100 |
Maximum | 100 | 100 | 1000 | 1000 |