Skip to main content
. 2006 Jul 22;36(Suppl 2):126–135. doi: 10.1007/s00247-006-0220-4

Table 3.

Suggested EEIR values for pediatric fluoroscopic imaging equipment. Values recommended for a 23-cm FoV image receptor, ‘normal–exposure rate, 30 pulses per second during fluoroscopy, at 80 kVp, with standard total filtration of 2.5 mm aluminum. Numbers in the table are in units of air kerma in nGy per image

Operational mode EEIR range
‘Low–fluoroscopy 6–.5
‘Medium–fluoroscopy 12–7
‘High–fluoroscopy 24–4
Digital angiography 450–00
Digital subtraction angiography 4,500–,000
Cardiac digital 90–30