Skip to main content
. Author manuscript; available in PMC: 2009 Apr 1.
Published in final edited form as: J Liq Chromatogr Relat Technol. 2008 Jan 1;31(9):1346–1357. doi: 10.1080/10826070802019913

Table 2.

Partition efficiency and stationary phase retention of polar two-phase solvent system in spiral HSCCC

Two-phase solvent system: 1-BuOH/AcOH/H20 (4:1:5)

Sample: trp-trp (1 mg), trp-tyr (1 mg), leu-tyr (4 mg), val-tyr (4 mg), tyr-gly (4 mg) in 1 ml solvent (UP 0.4 ml + LP 0.6 ml)

Column: Sprial tube assembly, 1.6 mm ID, cross-pressed, 4 radial grooves compressed, 12 layers, 100 ml capacity, embedded in PEG3350

Elution mode Flow rate (ml/min)
2 3 5
TP RS Sf% TP RS Sf% TP RS SF%
LIH 414/292/195/194/- 2.93 50 676/415/198/210/- 2.48 44 -/-/-/189/145 - 29
LIT 784/442/299/392/- 4.23 52 1810/661/360/391/- 3.61 43 -/-/1079/406/344 - 14
UOT 522/331/252/290/191 2.66 39 345/333/210/256/166 2.12 30 -/-/-/-/169 - 19
UOH 1024/784/351/398/623 3.98 70 900/816/492/622/442 3.69 50 -/1764/1239/917/375 3.48 32

TP: theoretical plate number (1st/2nd/3rd/4th/5th peaks); Rs: peak resolution between the 2nd and the 4th peaks; Sf: stationary phase retention - : measurement was not made.