Table 2.

Dose discrepancy, calculated by subtracting full 4D summation dose from the AVG-CT dose, in cGy. The percentage was calculated by dividing the dose difference by the 4D summation dose and converting to a percentage for each numerical motion phantom. GTV D99 and D1 represent the dose to 99% and 1% of the GTV, while MLD stands for mean lung dose.

S-I Amplitude (cm) Dose discrepancy (cGy)
GTVMin GTVMean GTV D99 GTV D1 MLD
2 0.27 0.35 0.80 0.45 0.25
(0.30%) (0.37%) (0.88%) (0.47%) (0.25%)
3 0.56 0.55 0.70 0.70 0.39
(0.63%) (0.58%) (0.77%) (0.73%) (1.01%)
4 1.74 0.29 1.20 0.60 0.39
(1.99%) (0.31%) (1.34%) (0.63%) (1.01%)
2.7 cm diaphragm-tumor distance 0.81 0.03 0.75 0.10 0.22
(0.89%) (0.03%) (0.82%) (0.10%) (0.58%)
4.7 cm diaphragm-tumor distance 0.19 0.21 0.40 0.45 0.63
(0.22%) (0.23%) (0.44%) (0.47%) (1.61%)