. 2008 Nov 24;35(12):5944–5953. doi: 10.1118/1.3013563

Table 3.

Comparison of registration results for the phantom and a sample patient dataset for our in-house B-spline-based method, using the same parameter settings. The results for error in registration of the inhale state to other deformation states of the phantom (1 and 2 cm compressions) are also shown.1

    RL (mm) AP (mm) SI (mm) 3D (mm)
Sample patient d¯ −1.0 −0.8 −0.5 2.2
  σ 1.2 2.1 1.8 2.4
  dmax −3.3 −5.1 −3.3 6.9
Phantom 3 cm diaphragm compression d¯ 0.0 −0.1 −0.2 0.8
σ 0.4 0.4 0.8 0.6
  dmax −1.7 −1.3 −3.1 3.1
Phantom 2 cm diaphragm compression d¯ −0.2 −0.2 −0.2 0.7
σ 0.3 0.4 0.6 0.4
  dmax −1.5 −1.4 −2.3 2.3
Phantom 1 cm diaphragm compression d¯ −0.3 −0.2 −0.1 0.7
σ 0.3 0.3 0.7 0.5
  dmax −1.1 −1.0 −2.8 2.8
a

d¯ is mean error; σ is standard deviation of error; and dmax is maximum error in estimation of marker position.