. 2008 Nov 20;35(12):5799–5820. doi: 10.1118/1.3013555

Table 2.

Prospective clinical trial of commercial CADe systems for screening mammography. These studies compared the statistical results by a group of radiologists over two periods of time before and after CADe was implemented in the practice. The patient cohorts were different and the radiologists may not be the same.

Investigators Number ofexams(unaided) Number ofexams(aided) Change incancerdetection rate (%) Change inrecallrate (%)
Gur et al. (Ref. 393) 56 432 59 139 +1.7 +0.1
      (24 radiologists)
  82 129 44 629 −3.3 −4.9
      (7 high-volume radiologists)
Feig et al. (Ref. 394) 11 803 21 639 +19.7 +14.1
      (17 low-volume radiologistsin Guret al.study)
Cupples et al. (Ref. 395) 7872 19 402 +16.1 +8.1
Fenton et al. (Ref. 396)1 398 159 31 186 +4.5 +30.7
Gromet et al. (Ref. 397) 112 413 118 808 +11.1 +3.9
a

From a survey study of 43 facilities, seven of which implemented CAD during the study period between 1998 and 2002.