Table 2 Temperature outcomes for preterm infants (23–27 weeks) transported via incubator or radiant warmer (median values with interquartile ranges are shown where appropriate).

Incubator Radiant Warmer p value
Number 30 32
Below 36.5°C (%) 10 (33) 7 (22) 0.4
Above 37.5°C (%) 2 (6.7) 1 (3.1) 0.60
Outside range (36.5–37.5)°C (%) 12 (40) 8 (25) 0.28
Below 36.0°C (%) 6 (20) 3 (9.4) 0.29
Outside range (36.0–37.5)°C (%) 8 (26.7) 4 (12.5) 0.20
Admission temperature (IQ range)°C 36.8 (36.2–37.3) 36.8 (36.5–37.1) 0.81
Time to admission (min) 17 (14–18.3) 14.5 (11–19.8) 0.21
Coefficient of variation 0.46 0.28 0.51
Maternal temperature (°C) 36.7 (36.2–37.2) 36.9 (36.4–37.2)