TABLE 1.
Step | Problem | Possible reason(s) | Solution |
---|---|---|---|
Steps 1-34 | Incomplete hardening of the resin tissue blocks | Incomplete tissue dehydration and/or resin infiltration | Check the temperature of reagents for each step and extend timing as needed |
Inadequate exposure of blocks to ultraviolet light or failure to keep the cryochamber temperature within correct range | Check that the ultraviolet light source of the chamber is intact and increase the frequency of temperature monitoring to maintain a range between -10 °C to -20 °C for 24-48 h | ||
Nonspecific gold label (i.e., over non-tissue regions such as the lumen of a vessel or capillary) | Blocking serum is inappropriate for the antibody used | If appropriate, adjust the blocking serum | |
Primary antibody is inappropriate for the tissue | Check that the antibody is appropriate, and the expiry date and dilution of the antibody and gold solution used | ||
Gold label is low or absent | Antibody failed to bind antigenic sites | Run a dilution experiment to establish the optimal antibody concentration | |
Gold solution failed to bind antibody-labeled sites | Complete a run on tissue that is a known positive control | ||
Background debris (dirt) is high | Contaminated solutions | Add additional rinsing steps and ensure that all solutions are filtered immediately before use | |
Steps 35-56 | Cell sample does not become red-transparent | Incomplete lysis of RBCs | Centrifuge at 300g at 4 °C (with brake) for 5 min to remove the supernatant and repeat the lysing step for 1-2 min at RT |
Fluorescence signal is poor/inadequate | Instrument malfunction (i.e., a weak excitation signal is produced) | Use an alternate machine to determine if the problem is related to the hardware or reagents | |
Antibody/reagents are inappropriate or used at wrong concentration | If an instrument is malfunctional, use another cytometer; if not, check and repeat the experimental and staining procedures | ||
Inaccurate/shifting compensation matrix indicated by variable readings for nonspecific IgG-staining tubes for blood or tissue | Compensation matrix settings have changed | Run single-color controls, or alternating-color controls, to reset and confirm that the compensation matrix is accurate |
● TIMING
Steps 1-8: 48 h to dehydrate and infiltrate lung tissue blocks (36-48 per rodent group) with resin
Steps 9-13: 58 h to polymerize tissue blocks in Unicryl resin, label and archive them
Steps 14-16: 8 h to cut 3-4 tissue blocks, that is, 2 h to rough-cut 1 tissue block, and prepare 1-μm-thick sections on glass slides and (approximately 15-20) 90-nm-thick sections on Formvar-coated grids
Steps 17-20: 16 h to treat (approximately 15-20) 90-nm-thick sections with primary antibody
Steps 21-26: ~2-3 h to label 15-20 grids/sections with pA-AU; 3 h to treat approximately 15-20 grids/sections with silver halide (Box 1)
Steps 27-31: 4 h to enhance electron-density of cell membranes and organelles in 15-20 grids/sections
Steps 32-34: ~2-3 d to view multiple grids/sections by TEM, collect overviews and determine the appropriate analysis. Allow an additional ~3 h to collect 100 images for qualitative or quantitative analysis, and 2 min per cell to quantify antigenic sites using the macro program
Steps 35-40: ~2 h to prepare cell suspensions from blood
Steps 41-48: ~2 h to prepare cell suspensions from lung tissue
Steps 49-56: 5 h to immunostain cells and 30 min to analyze six tubes by flow cytometry after a compensation matrix has been established