Skip to main content
. Author manuscript; available in PMC: 2010 May 8.
Published in final edited form as: J Chromatogr A. 2008 Nov 13;1216(19):4193–4200. doi: 10.1016/j.chroma.2008.10.126

Table 5.

Separation of proteins by a spiral tube assembly with cross-pressed and 4-way compressed 1.35 mm ID PTFE tube with 15 spiral layers and a total capacity of 85 ml.

Flow 0.5 ml/min 1 ml/min 2 ml/min
rpm TP Rs SF TP Rs SF TP Rs SF
600 43/54(35/28* 1.62 (1.04) 55.9(63.0) 72/50(56/30) 1.04(1.13) 39.4(54.3) 144/46(46/25) 0.82(0.58) 21.2(19.7)
800 96/66(27/32) 1.75 (1.27) 63.2(76.6) 86/80(44/44) 1.50(1.24) 50.0(61.5) 144/104(128/66) 1.18(1.11) 35.3(44.2)
1000 51/76(40/91) 1.90 (1.61) 70.6(76.0) 85/104(77/90) 1.88 (1/78) 55.9(71.2) 228/275(88/138) 2.18(1.81) 41.2(56.7)
1200 68/91 2.16 70.6 144/168(92/122) 2.61 (2.29) 67.1(71.2) 455/249(133/191) 2.64(2.06) 47.1(59.6)

TP: theoretical plate numbers for the first / second peaks; **Rs: Peak resolution; SF: % retention of the stationary phase; elution modea: L-I-T

*

Figures in parentheses indicate the date from 4-way compressed plain tube with 15 layers and 120 ml capacity.