. 2009 Mar 4;113(20):4853–4855. doi: 10.1182/blood-2008-12-191643

Table 1.

Comparison of Bland-Altman statistics by percentage difference and standard methods for different R2-MRI and R2*-MRI methods

Comparison method / MRI method Bias SD 95% confidence interval
Percentage difference method
    R2*-MRI (using robust regression analysis), % 3 21 −39 to 45
    R2*-MRI (Wood et al3; using least squares regression analysis), % 1 23 −46 to 44
    R2-MRI (St Pierre et al2; using nonlinear regression algorithms), % −3 27 −56 to 50
Standard method
    R2*-MRI (Altman and Bland6,7; using robust regression analysis), mg Fe/g 0.0757 1.84 −3.60 to 3.76

The Bland-Altman percentage difference method8 was calculated using the formula (HIC [R2*] − HIC [biopsy])/([HIC (R2*) + HIC (biopsy)]/2).