TABLE 2.

Correlation between results of Architect assay and those of Vidas, Enzygnost, and AxSYM CMV IgM assays with 503 routine samples

Architect assay result No. of samples with the indicated result by the following assay:
Vidas assay
Enzygnost assay
AxSYM assay
Negative Gray zone Positive Negative Gray zone Positive Negative Gray zone Positive
Negative 457 3 0 432 25 3 434 10 16
Gray zone 2 0 0 1 0 1 2 0 0
Positive 8 3 31 1 8 33 7 3 32