TABLE 4.

Hypothetical matrix relating lysate batch size to the potential for discard of freshly reconstituted master mix under current BD GeneOhm MRSA assay specifications

Lysate batch size (size with controls included) No. (%) of unused aliquots following reconstitutiona
8 aliquots/vial 9 aliquots/vial 10 aliquots/vial
1 (3) 5 (62.5) 6 (66.7) 7 (70.0)
2 (4) 4 (50.0) 5 (55.6) 6 (60.0)
3 (5) 3 (37.5) 4 (44.4) 5 (50.0)
4 (6) 2 (25.0) 3 (33.3) 4 (40.0)
5 (7) 1 (12.5) 2 (22.2) 3 (30.0)
6 (8) 0 (0.0) 1 (11.1) 2 (20.0)
7 (9) 7 (43.8) 0 (0.0) 1 (10.0)
8 (10) 6 (37.5) 8 (44.4) 0 (0.0)
9 (11) 5 (31.3) 7 (38.9) 9 (45.0)
10 (12) 4 (25.0) 6 (33.3) 8 (40.0)
11 (13) 3 (18.8) 5 (27.8) 7 (35.0)
12 (14) 2 (12.5) 4 (22.2) 6 (30.0)
13 (15) 1 (6.3) 3 (16.7) 5 (25.0)
14 (16) 0 (0.0) 2 (11.1) 4 (20.0)
a

Mandated controls were added to the batch size prior to the calculation of unused reaction tubes. The matrix assumes the presence of one 16-thermocycler block in a laboratory. The mean percentages of unused aliquots per random distribution batch size were 24.4, 29.6, and 33.5%, respectively, for 8, 9, and 10 aliquots per master mix vial.